2.4 Error messages
Temperature module TED8xxx / page 41
2.4 Error messages
Error messages are shown in the bottom line of the display not regarding, if you are
in the main menu or in the channel menu.
If an error occurs, the display shows for example:
CH3 T
S
= 22.000 °C
TE on T
A
= 22.000 °C
Th lo Twin= 5.60 °C
<CH3 Sens.fail> ok
Channel causing the error
Error message
ok to accept
Possible error messages for a TED8xxx module are:
Sens.fail
Wiring to the sensor has opened during operation
No Sensor
No or wrong sensor connected when trying to switch on
OPEN
Wiring to the Peltiér has opened during operation
OTP
Module is too hot. Operation is possible again after cooling
down
TEMP PROT
Temperature has left the window (e.g. the laser has been
switched off).
NOT IF TEC ON
Some parameters may not be changed with the
temperature controller switched on.
OVERTEMP
The module is too hot and can not be switched on. Wait
some time for cooling down.