3.7 Error messages of the TED350
TED350 / page 51
[
303, "Key emulation error"
]
Category:
Device
Error
Possible reason:
Internal communication problem. Please contact
Thorlabs
.
[
400, "Too many errors"
]
Category:
Device
Error
Possible reason:
Error queue overflow (32 errors max).
[
410, "Query interrupted"
]
Category:
Query
Error
Possible reason: More than one query sent to the TED350 before the read
command.
[
420, "Query unterminated"
]
Category:
Query
Error
Possible reason:
There is no data in the output buffer.
[
500, "IEEE488 receive buffer overflow"
]
Category:
Device
Error
Possible reason: The string sent to the TED350 was too long for the IEEE488
receive buffer (250 char max).
3.7.2 TED350 operation error messages
[
1103, "Over temperature"
]
Category:
Execution
Error
Possible reason:
Internal temperature too high. Wait until the TED350 has cooled
down. Maintain proper air flow.
[
1104,
"Wrong or no sensor"
]
Category:
Execution
Error
Possible reason:
The TED350 detected no sensor at the output jack.
[
1106,
"Wrong command for this sensor"
]
Category:
Execution
Error
Possible reason:
The released command does not fit to the connected sensor.