2.1 Manual operation
TED350 / page 21
2.1.4 The output ANALOG OUT
This output offers a voltage in strict correlation to the difference between actual and
set temperature, scaled to the set temperature window.
The upper border of the temperature window 10 V output and the lower
border -10 V.
Example:
If the set temperature is 20 °C and the temperature window is set to 5 °C an actual
temperature of
15 °C leads to -10V at the output
20 °C leads to 0 V at the output
25 °C leads to +10 V at the output
.
.
.
2.1.5 Switching the output ON and OFF
With the button TEC ON the TEC current is switched on or off. The LED inside this
button indicates the state of the output.
2.1.6 Setting the PID shares
(Refer to chapter 2.2, “Optimization of temperature control loops“ starting on page
24)
2.1.7 WIN OUT
This jack can be used to supervise the temperature of a laser diode when using a
Thorlabs
LDC340 laser diode driver. Simply connect the corresponding jack of the
LDC340 to this output of the TED350. Further details can be found in the operation
manual of the LDC340.