1.7 Connecting components
TED350 / page 16
If the voltage of the TEC element shall be measured and displayed two additional
cables must be connected directly from the two poles of the TEC element to pin 2
and 9 of the jack. (4-point measurement).
Pin 2 and pin 9 may also be connected directly in the plug at the jack (i.e. pin 2 with
pin 5,6,7 and pin 9 with pin 13,14,15) but this leads to a measurement error due to
the voltage drop in the lines to the TEC element. Then a slightly too high voltage is
displayed.
13
15
14
-
7
5
6
+
9
2
Figure 9
Connecting the TEC element
Attention
A TEC element connected with wrong polarity may lead to thermal
runaway and destruction of the connected components.