2.2 Optimization of temperature control loops
TED350 / page 28
3 Communication with a control computer (TED350-IEEE)
NOTE
The following operating elements are still working in remote mode:
•
The display
•
The display selection
•
all potentiometers except the main dial knob
NOTE
All analog values are read and written in SI units, i.e. A (not mA), W (not
mW) etc. Letters may be written in small or capital letters.
Attention
Before programming a TED350 first set the limit value of the TEC
current I
TEC LIM
(hardware limit) for the used TEC element with a
screwdriver.
The corresponding potentiometer is marked ADJ and located next to
the LED I
TEC LIM
at the front of the TED350.
The value I
TEC LIM
is constantly measured and can be checked if the
LED I
TEC LIM
is selecting the display of the limit current.