3.4 System command group
TED350 / page 40
3.4 System command group
3.4.1 Answer mode
Programming:
Syntax:
"
:SYST:ANSW FULL
"
"
:SYST:ANSW VALUE
"
Default Value: FULL
Description: When switched to
"
VALUE
"
the TED350 sends only the requested
parameter without designator.
Example:
When requesting the actual TEC current with
"
:ITE:ACT?
"
the
TED350 will only send
[5.123456E-02]
instead of
[:ITE:ACT 5.123456E-02]
. This is
not
according to the
IEEE488.2 standard but useful if you want to increase speed.
Reading:
Syntax:
"
:SYST:ANSW?
"
Response:
[:SYST:ANSW FULL]
[:SYST:ANSW
VALUE]
Description: Queries the answer mode.
3.4.2 Error-LED mode
Programming:
Syntax:
"
:SYST:ERRLED OFF
"
"
:SYST:ERRLED ON
"
Default Value: OFF
Description: When switched to
"
ON
"
the interface ERRor LED lights as long as
there are errors in the error queue. When switched to
"
OFF
"
the in-
terface ERRor LED is dark.
Reading:
Syntax:
"
:SYST:ERRLED?
"
Response:
[:SYST:ERRLED OFF]
[:SYST:ERRLED
ON]
Description: Queries the Error-LED mode.