6.2.5 Save Settings
Command
Explanation
Response Example
"M"
Saves the setup. The actual parameters that have been set
using the commands
T, W, L, d, G, O, P, I, D, C and S, are saved
to the nonvolatile memory.
6.2.6 Factory Default Settings
Parameter
Explanation
Factory Default
L
TEC current limit
1000 mA
T
Temperature set value
25 °C
W
Temperature window
1000 mK
d
Temperature window delay
10 sec
C
Cycle time PID loop
50 msec
P
P share PID loop
1000 mA/K
I
I share PID loop
200 mA/(K*sec)
D
D share PID loop
100 (mA*sec)/K
G
Critical PID loop gain (Loop test)
2000 mA/K
O
Critical PID loop oscillation period (Loop test)
2000 msec
S
Value of the Safety Bitmask
255