© 2016 Thorlabs GmbH
14
MTD415T
MTD415T Data Sheet Rev. 1.0
6.2.3 Temperature
Command
Explanation
Response Example
Programming
"Tx"
Sets the set temperature to
x
*)
Value range
x
:
5000
to
45000
[10
-3
°C]
Reading
"T?"
Reads the set temperature
[x<LF>]
Value range
x
:
5000
to
45000
[10
-3
°C]
"Te?"
Reads the actual temperature
[x<LF>]
Command
Explanation
Response Example
Programming
"Wx"
Sets the set temperature window to
x
*)
Value range
x
:
1
to
32768
[mK]
Reading
"W?"
Reads the temperature window
[x<LF>]
[mK]
Command
Explanation
Response Example
Programming
"dx"
Sets the delay time between reaching the temperature window
and activating the
Status
output pin to
x
*)
Value range
x
:
1
to
32768
[sec]
Reading
"d?"
Reads the delay time between reaching the temperature
window and activating the
Status
output pin
[x<LF>]
[sec]
*) This set value remains during the actual power-on cycle. Use the "
M
" command to save it to the non-volatile
memory.