![THORLABS MNA601/IR Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/thorlabs/mna601-ir/mna601-ir_user-manual_1110707030.webp)
30
HA0118T Rev 6 Feb 2016
Chapter 3
AC Strain Gauge
- The AC Strain Gauge mode refers to the use of AC excited
strain gauge feedback signals as generated by the complete range of Thorlabs
piezo actuators and piezo equipped multi axis stages. All versions of the APT
NanoTrak electronics support this feedback mode.
DC 0-10V
- On APT modular hardware (and build 2 versions of benchtop APT
hardware) it is possible to set the feedback signal type to DC 0 – 10V. This
configures the feedback circuit to accept a differential 0-10V DC feedback signal
in order to control position output. This latter feedback mode is available for use
with third party piezo positioning systems capable of generating a standard
voltage position feedback signal.
3.6
Programmed NanoTrak Operation
3.6.1
A Typical Alignment using the NanoTrak
1) In the NanoTrak GUI, set the ‘Scan Circle Diameter’
control to approximately mid-
position.
2) Press the ‘Latch’ button to select Latch mode and check that the Latch indicator
LED is lit.
3) Click and drag to centre the scan circle in the display.
4) Press the ‘Auto’ button to select ‘Auto’ mode. Ensure that the ‘Auto’ LED is lit.
5) Adjust the actuator on the positioning stage such that the fibers are spaced well
apart, with zero power transmission between them. (Alternatively, switch off or
obscure the optical power source). Make a note of the bar graph displays of
relative power input and range. Since the optical source has been removed, only
background noise is registered.
6) Use the actuators to bring the fiber tips together, until the display shows a
significant increase indicating that the power has risen above the background
level.
7) Press the ‘Track’ button to select Track mode. Check that the ‘Tracking’ indicator
is lit and that the circle moves towards the position of maximum power (center
screen). The NanoTrak has now positioned the fiber to give the maximum optical
power transmission.
Note
On APT modular hardware (and build 2 versions of benchtop APT
hardware), the function of the pin connections on the 9-way D-type
feedback connector is dependent on the feedback mode selected. Refer
to the pin-out tables in Section 4.2.3. for details.
Note
Familiarize yourself with NanoTrak operation from the GUI panel before
attempting programmed operation from the Controller PC.