![THORLABS MNA601/IR Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/thorlabs/mna601-ir/mna601-ir_user-manual_1110707025.webp)
25
NanoTrak Autoalignment Controller
If
dB
is selected, readings are displayed or returned in decibels. The signal is
first converted to watts according to the calibration factor specified in the A/W
Calibration parameter, and then into dB.
A/W Calibration
- the Amp:Watt calibration factor used when Watts or dBs is
specified in the PIN Units Mode parameter.
BNC Reading Units
Units Mode
-.this parameter specifies the units relating to the input signal
measured by the NanoTrak, and is applicable only if a BNC input source has been
specified.
If
Voltage
is selected, readings are displayed or returned in Volts, i.e. the raw
input signal to the BNC connector.
If
% Full Range
is selected, readings are displayed or returned as a
percentage of full scale input, e.g. a 1V signal into a 2V input range BNC will
display a value of 50%.
If
User Calib
is selected, readings are displayed or returned as a calibrated
value using the Voltage Calibration parameter, e.g. if Voltage Calibration has
been set to '20', a 1V signal into a 2V input range BNC will display a value of
10.
Voltage Calibration
- the calibration factor used when User Calib is selected in the
BNC Units Mode parameter.
Channel Outputs
- these parameters allow the horizontal and vertical channels of the
NanoTrak to be enabled/disabled.
Open/Closed Loop -
Typically, the NanoTrak is operated in Open Loop mode
because the piezo amplifiers have a higher bandwidth operation in open loop, thereby
ensuring best bandwidth performance. If the NanoTrak is latched, the piezo actuator
position is best maintained if the drift is minimized. This is achieved by placing the
piezo channels into Closed Loop mode.
When in closed-loop mode, position is maintained by a feedback signal from the piezo
actuator strain gauge or by a DC feedback signal derived externally.
Note
. This is only
possible when using actuators equipped with position sensing.
Note
NanoTrak channels should not normally be disabled.