19
NanoTrak Autoalignment Controller
If ‘dB’ is selected, readings are displayed or returned in dBs. The signal is first
converted to Watts according to the calibration factor, and then into dB.
Consequently, this conversion is also wavelength dependent.
If a BNC input is connected (input source is set to a BNC mode using the SetInputSrc
method), the display can be configured to display 'V', '%' or 'User'.
If 'V' is selected, readings are displayed or returned in Volts, i.e. the raw input
signal to the BNC connector.
If '%' is selected, readings are displayed or returned as a percentage of full scale
input, e.g. a 1V signal into a 2V input range BNC will display a value of 50%.
If 'User' is selected, readings are displayed or returned as a calibrated value using
a calibration factor set using the SetInputSrc method (or the Voltage calibration
parameter on the Input/Output tab of the Settings panel), e.g. if the calibration
factor has been set to '20', a 1V signal into a 2V input range BNC will display a
value of 10.
Display Averaging
- displays the averaging level being applied to the power readings
(and the relative display bar) displayed on the GUI. Settings of 'low', 'medium' and
'high' resulting in averaging numbers of 5, 10 and 20 repectively. At the default display
update rate of 10Hz, this translates to 0.5, 1 and 2 second time constants
respectively. This setting is provided as a convenient way of smoothing readings
displayed on the GUI only, and does not affect in any way those same readings when
returned by ActiveX methods.
To change the setting, click the arrow buttons.
Scan Circle Diameter control
- used to adjust the diameter of the scan circle, when
operating in 'Manual' adjustment mode. The current diameter is displayed in the
'Settings' window. Note. Use of this control will automatically switch the circle
diameter adjustment mode to 'Manual'.
Relative SignalPower -
The max power located in the alignment. Bar display
showing a convenient visual indication of the feedback signal level. When the input
source is set to TIA (PIN), the bar display indicates the power level with respect to the
currently applied range. E.g. if the TIA range is 1micro, then a bar displaying half
length equates to a power level of 0.5micro. When the input source is set to BNC
(voltage), then bar display indicates the voltage being measured. E.g. for an input
voltage range of 2volts, a bar displaying half length equates to 1volt input signal.
Track
- push button control used to set the NanoTrak to 'Track' mode, using both
horizontal and vertical axes to set a circular scan pattern. When tracking, the
NanoTrak detects any drop in signal strength resulting from misalignment of the input
and output devices, and makes positional adjustments to maintain the maximum. The
LED in the button is lit when selected.
Track Horz
- push button control used to set the NanoTrak to track using a horizontal
line scan pattern. The LED in the button is lit when selected.
Track Vert
- push button control used to set the NanoTrak to track using a vertical
line scan pattern. The LED in the button is lit when selected.