Chapter 3
NanoTrak Screen tab
Fig. 3.4 NanoTrak Settings panel - NanoTrak Screen Tab
Sometimes, it may be useful to explore the area of interest by manually
repositioning the two piezos driven by the NanoTrak. If desired, the NanoTrak
circle can be positioned by using the mouse. This feature allows manual
repositioning of the piezos, e.g. to enable investigation of the region of interest to
look for side peaks, artifacts etc.
The circle can be moved either by click and drag, or by moving an incremental
distances per mouse click.
Note. The circle can be moved in both Latch and Tracking modes.
Relative Circle Move Increments
- When the circle is positioned by a mouse click,
the incremental distance can be set to NT units or microns. If set to NT Units, the
increments can be set from 0.01 to 2 NT units.
If set to microns, the limits are dependent upon the travel of the piezo actuator,
e.g. for a 20µm piezo travel, 1 NT unit = 2µm, therefore the limits are 0.02 to 4µm.
Display Units
- The display at the bottom of the NanoTrak screen shows the circle
diameter and the horizontal and vertical co-ordinates of the circle position.
Fig. 3.5 NanoTrak screen - circle coordinates
This data can be set to display in microns or NanoTrak units.