1.3 Features
ITC5000
/ page 3
1.3 Features
1.3.1 Safety measures for the laser diode and the TEC element
To protect the connected laser diodes and the TEC elements the combi controller
ITC5xxx contains the following protection circuits:
•
Softstart when switching on the laser diode current
Protection against capacitive and inductive parasitic elements (switching peaks).
•
Software limit for the injection current and the TEC element in all operating
modes
Protection against thermal destruction of the devices.
•
Interruption control of the connection cable to the laser diode (interlock)
Protection against accidental operation.
•
Contact protection of the laser diode (open circuit)
Protection against cable damage or bad contact.
•
Electronic short-circuit switch for the laser diode
Protection against static discharge when touching the switched off laser.
•
Protection of the sensor
Protection against the use of an wrong temperature sensor / protection against line
interruption to the temperature sensor.
•
Separate on and off function for each module
Protection against operating errors.
•
Separate over-temperature protection for each module
Protection against thermal failure of the module.
•
Control LED for the laser and/or TEC current active