1.7 Connecting components
ITC5000
/ page 15
1.7.2 Connecting the Laser diode
Connect laser and monitor diode to the connector of the ITC5xxx.
The wires for voltage measurement of the laser diode (pin 12 and pin 9) should be
connected as close as possible to the laser diode to avoid measurement errors.
The ground conductor of the monitor diode (pin 2) may be connected to the ground
conductor of the laser diode (pin 3). If this is necessary (e.g. with laser diodes with
integrated monitor diode and shared ground connector) the ground conductors
should be connected as close as possible to the laser diode to avoid measurement
errors when measuring the monitor diode current.
The following pin assignments of the output jack are possible:
(shown with voltage measurement)
Figure 2
Connecting the Laserdiode
+
LDA (11)
ULD+ (12)
LGND (3)
ULD- (9)
LDK (10)
+
LDA (11)
ULD+ (12)
LGND (3)
ULD- (9)
LDK (10)
Cathode Grounded (CG)
Anode grounded (AG)
Содержание ITC5000
Страница 1: ...Operation Manual Thorlabs TXP Series Laser Current and Temperature Controller ITC5000 2004...
Страница 2: ...Version 2 10 Date 11 Oct 04 Copyright 2004 Thorlabs GmbH...
Страница 6: ......
Страница 48: ......