6.1. Changing the Input Fuses
If for some reason you need to replace an open fuse in the base unit, you must perform the following procedure:
Remove the AC input cable that may be connected to the unit.
Slide open the cover of the fuse holder located at the rear panel of the base unit as shown in Figure 23.
Remove the broken fuse(s) and install the appropriate replacement fuse(s) for the base unit. Use only
IS 1 A 250 VAC Type T 5x20 mm style fuses (IEC 60127-2/III, low breaking capacity, slow blow).
Slide the fuse cover closed.
Figure 63 Fuse Cover on Base Unit Rear Panel
100-240 VAC 50-60 Hz
45W MAX.
FUSE 2x 1A/250V
TYPE T 250