SD-OCT Base Unit
Chapter 5: Imaging Artifacts
Rev E, November 7, 2018
Page 39
When operating with a wedge, the image will be tilted in the direction of the wedge angle. When scanning in the
orthogonal direction, no tilt occurs.
5.2.
Wrong Reference Intensity Setting
The OCT image is created by interferometry as shown in Figure 17. For good image acquisition, the intensity
of the reference light needs to be well above noise level and well below saturation. In the OCT software, a
colored bar indicates if the reference intensity is set properly. For adjustment of the reference intensity, please
refer to chapter 3.2.4.
Figure 49 OCT Image Acquired with Low (A), High (B), and Good (C) Reference Intensity Settings
With low reference intensity (see picture Figure 49A), the image becomes very noisy and autocorrelation is
strong compared to the signal intended (frequent issue when doing thickness measurement of reflecting films).
High reference intensity (see picture Figure 49B), causes saturation and loss of information.
A
B
C