18. VOLTAGE OF THE ACCUMULATOR / Status of Power Supply :
OK
:
voltage >11.5 V
!!!
voltage 10.6 ... 11.5 V
Low
:
voltage <10.5 V loading/replacing of the accumulator !
Display of the measured voltage of the accumulator.
Accumulator: OK
12.5 V
Mains AC: 1
Line 3 displays the state of power supply (1: mains voltage present, 0: no mains voltage).
Remark:
Analogue measurements at a recorded voltage of below 9V are not accurate!
A discharge of the accumulator below 10,5 V should be avoided, as, firstly, no
considerable capacity is available any more, and secondly the life time of the
accumulator is reduced considerably ! Please replace or charge the accumulator,
if "!!!" is displayed. In order to protect the accumulator against further discharge
the query of the sensors is interrupted at a voltage of 10.5 V; thus the current
consumption is minimized . Then, the voltage is checked every 5 minutes; when it
is higher than 11,0 V the normal measuring routine is continued (ref. to chap.4).
19. DCF77 RECEIVING CONTROL:
26 - 56
021579/06/09
DCF77 Test: ss nn
D:1.0s L:10 0
DCF77 Test: --
DCF-antenna on
DCF-antenna off
For more information refer to chap.
20. DCF77 SYNCHRONISATION
*DCF: 1 !!.!!.!!
ffff n !!:!!:!!
!!.!!.!! !!:!! 0
DCF-antenna on
DCF-antenna off
*DCF77: 0
!!.!!.!! !!:!! 0
For more information refer to chap.