60 - 66
021341/07/11
Minimum Intensity
(depending on precipitation type)
< 0,005 mm/h (drizzle)
Maximum Intensity
> 250 mm/h
Visibility MOR in precipitation
0 … 99999 m
Radar reflectivity
-9.9 … 99.9 dBZ
Disdrometer classes
440 classes (22 diameter * 20 speed)
Option Measuring Inputs: 5.4110.1x.xxx
Temperature (Pt100) (not available for variant
5.4110.
11
.xxx)
Meas. range: -40... 70°C
Accuracy: ±0.1°C (measuring input)
Relative humidity (0-1V)
Meas. range: 0 ... 100 % r. h.
Accuracy: ±0.1% (meas. range)
Wind Speed (Thies Compact, e.g. 4.3519.00.000) Meas. range: 0 ... 60 m/s
Accuracy: ±0.1m/s (measuring input)
Wind direction (Thies Compact, e.g.
4.3129.00.000)
Meas. range: 0 ... 360°
Accessories (Optional):
Visualisation software LPM View (9.1700.99.000)
For Windows operating systems (Win98(SE), WinMe,
Win2000, XP).
Instrument Support (4.3187.60.200)
For the vibration-reduced operation of the LPM at
2 m height on an available concrete foundation,
provided by the customer.
Tube length: 2 m
Tube diameter: 60 mm
Weight: 30 Kg
Material: Steel, hot-dip galvanized
Alteration of technical data reserved
*
Supply must be potential-free, or the negative voltage is on the same potential as the
housing of the sensor (protective earth). This UPS-supply (“uninterruptible power supply”)
allows a restricted operation (no heating) of the sensor whenever the normal supply falls
out.
**
Applies for liquid and solid particles like rain, pellets/graupel and hail. For snow flakes by
reason of the less optical extinction a greater maximum diameter can be assumed.
***
Visibility in the precipitation: Only atmospheric hazes caused by big hydro-meteors (for ex.
drizzle, rain, snow) are acquired.