background image

P a g e

 | 

 

17/03/11 

ELECTRICAL - POWER LEAD 
 

A removable mains power lead with a three pin plug and right angle female IEC 
socket is supplied for connection between the male IEC socket located at the 
incubator (top left side) and a ten (10 Amp General Purpose Outlet (G.P.O.)

 

 

OVER TEMPERATURE INDICATION 

 

All Eurotherm 3216 controllers are fitted with an automatic over temperature alarm. In 
the event of a malfunction of the Thermoline product, the alarm will switch off all 
forms of heating if the actual temperature exceeds the set temperature by 2.0°C. This 
wil

l be indicated by the red “over temp” indicator and a scrolling message of ‘Over 

Temperature’ on the bottom display of the controller. 

 

OPERATION 
 

WARNING: 1. 

To avoid injury do not use flammable solvents in this 

cabinet, as fire or explosion may result. This cabinet contains 
components that may ignite such materials. 
 

WARNING: 2.

 Do not introduce water into this chamber as the 

excessive vapour may result in premature component failure.

 

 
 

OPERATING THE INCUBATOR 

 

1)  Ensure that the incubator is located as previously described, refer to the 

index, LOCATION, Page 2. 

 

 

2)  Position the mains switch to the ON position (reset) and immediately the 

circulating fan, if fitted, will start and the controller will go through a short 
initiation period, the two displays will then indicate the operating value at the 
top and the set value at the lower display.  

 
3)  The temperature controller is a Eurotherm 3216 microprocessor based 

instrument which has been factory configured for range, sensor type, and 
engineering parameters for optimum control.  
Limited access to the control is available to the operator, refer to the index, 

CALIBRATION

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Содержание TI-150F

Страница 1: ...Scientific Equipment Pty Ltd T A Thermoline Scientific ABN 80 000 859 129 10 12 Ross Place Wetherill Park N S W 2164 Australia P O Box 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 0...

Страница 2: ...Possibility of damage to the Equipment To maintain safe operation of this equipment be aware of the ALERT BOXES located throughout these operating instructions CONTENTS SAFETY INFORMATION 2 ALERT BOXE...

Страница 3: ...ssemble the components and inspect for damage Retain the packaging materials until the equipment has been thoroughly tested Notify the detail of any damage without delay to your supplier or Thermoline...

Страница 4: ...viewing products under evaluation whilst not disturbing the conditions is standard in each model Removable chrome plated open wire shelves are supplied with additional shelf spacing The desired opera...

Страница 5: ...han normal levels of cleanliness The external surface powder coated steel and the inner surface stainless steel may be wiped clean using a damp soft cloth CARING FOR STAINLESS STEEL Stainless steel is...

Страница 6: ...it breaker continues to trip it may indicate an electrical fault exists Switch Off the power and consult a qualified technician TEMPERATURE CONTROLLER SENSOR This is a Eurotherm 3216 PID control with...

Страница 7: ...use flammable solvents in this cabinet as fire or explosion may result This cabinet contains components that may ignite such materials WARNING 2 Do not introduce water into this chamber as the excessi...

Страница 8: ...the temperature ramps from one temperature at a set rate to a second temperature and then dwells at that temperature for a set period of time This ramp dwell program has 4 ramp dwell segments CHANGING...

Страница 9: ...our TGSP 2 40 C RMP 2 5 C hr Dwell at 40 C for 5 hours DWEL 2 5 Ramp from 40 C to 45 C in one hour TGSP 3 45 C RMP 3 5 C hr Dwell at 45 C for 5 hours DWEL 3 5 Ramp from 45 C to 50 C in one hour TGSP 4...

Страница 10: ...ters shown below TSP 1 Target set point 1 RMP 1 Ramp rate 1 Note with RMP 1 OFF the temperature will rise from ambient to 35 C as quickly as possible DWEL 1 Dwell time 1 3 As for segment 1 described a...

Страница 11: ...ur then dwell there for 5 hours 5 As for segment 3 described above use the Scroll key to select in turn the tsp 4 parameter Target set point 4 the RMP 4 parameter Ramp rate 4 and the DWEL 4 parameter...

Страница 12: ...tat This symbol refers to Timer status reset which means that the program or timer is in the reset mode and is not running The status can be Res Run or Hold leave the program in reset mode until after...

Страница 13: ...temperature displayed and temperature measured at the centre of the workspace For this reason the digital temperature controls have a parameter that can be used to correct the temperature displayed ca...

Страница 14: ...will show PV OFS Use the up or down keys to enter an offset of 1 0 C Use the Home key to return to the original display You will now notice that the 1 0 C calibration has been added to the incubator t...

Страница 15: ...warranty service is required Thermoline Scientific will endeavour to have an authorised service organisation carry out the work if there is an available agent however if it is decided to return the e...

Страница 16: ...ly by Any defect in material or workmanship of or any other defect whatsoever in or unsuitability for any purpose of the goods or any part of the goods or default or negligence on the part of Thermoli...

Отзывы: