EXPLANATION OF CONTROLS 
 

The controls are located in a control enclosure above the door. 
 
 

MAIN SWITCH 
 

This is a combination power switch and circuit breaker, when pressed this switch 
allows power to be supplied to the controls and circulating fan if applicable. 
Incorporated in the face of the 

switch is a ’MAINS ON’ indicating light. 

 

CAUTION: 

The equipment is fitted with a combination circuit 

breaker, mains switch. If the circuit breaker continues to trip, it 
may indicate an electrical fault exists. Switch Off the power and 
consult a qualified technician. 

 
 

TEMPERATURE CONTROLLER / SENSOR 
 

This is a Eurotherm 3216 PID control with digital temperature indication of both the 
set value and the operating value to 0.1°C. 
 
Engineering parameters are configured to provide stable temperature control with 
minimum over shoot and quick recovery times. 
 
Sensor type is a DIN standard platinum resistance bulb, RTD Pt100.

 

 

Maximum recommended temperature is 60°C.

 

 
 
 

 

                                  

 

 
 

 
 

 
 
 

OPERATING 

VALUE 

SCROLL 

BUTTON 

SET VALUE 

(SP) 

MAINS SWITCH 

(CIRCUIT 

BREAKER) 

DOWN 

ARROW

UP 

ARROW 

HOME 

BUTTON 

OVER 
TEMP 

LIGHT 

Содержание TI-150F

Страница 1: ...Scientific Equipment Pty Ltd T A Thermoline Scientific ABN 80 000 859 129 10 12 Ross Place Wetherill Park N S W 2164 Australia P O Box 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 0...

Страница 2: ...Possibility of damage to the Equipment To maintain safe operation of this equipment be aware of the ALERT BOXES located throughout these operating instructions CONTENTS SAFETY INFORMATION 2 ALERT BOXE...

Страница 3: ...ssemble the components and inspect for damage Retain the packaging materials until the equipment has been thoroughly tested Notify the detail of any damage without delay to your supplier or Thermoline...

Страница 4: ...viewing products under evaluation whilst not disturbing the conditions is standard in each model Removable chrome plated open wire shelves are supplied with additional shelf spacing The desired opera...

Страница 5: ...han normal levels of cleanliness The external surface powder coated steel and the inner surface stainless steel may be wiped clean using a damp soft cloth CARING FOR STAINLESS STEEL Stainless steel is...

Страница 6: ...it breaker continues to trip it may indicate an electrical fault exists Switch Off the power and consult a qualified technician TEMPERATURE CONTROLLER SENSOR This is a Eurotherm 3216 PID control with...

Страница 7: ...use flammable solvents in this cabinet as fire or explosion may result This cabinet contains components that may ignite such materials WARNING 2 Do not introduce water into this chamber as the excessi...

Страница 8: ...the temperature ramps from one temperature at a set rate to a second temperature and then dwells at that temperature for a set period of time This ramp dwell program has 4 ramp dwell segments CHANGING...

Страница 9: ...our TGSP 2 40 C RMP 2 5 C hr Dwell at 40 C for 5 hours DWEL 2 5 Ramp from 40 C to 45 C in one hour TGSP 3 45 C RMP 3 5 C hr Dwell at 45 C for 5 hours DWEL 3 5 Ramp from 45 C to 50 C in one hour TGSP 4...

Страница 10: ...ters shown below TSP 1 Target set point 1 RMP 1 Ramp rate 1 Note with RMP 1 OFF the temperature will rise from ambient to 35 C as quickly as possible DWEL 1 Dwell time 1 3 As for segment 1 described a...

Страница 11: ...ur then dwell there for 5 hours 5 As for segment 3 described above use the Scroll key to select in turn the tsp 4 parameter Target set point 4 the RMP 4 parameter Ramp rate 4 and the DWEL 4 parameter...

Страница 12: ...tat This symbol refers to Timer status reset which means that the program or timer is in the reset mode and is not running The status can be Res Run or Hold leave the program in reset mode until after...

Страница 13: ...temperature displayed and temperature measured at the centre of the workspace For this reason the digital temperature controls have a parameter that can be used to correct the temperature displayed ca...

Страница 14: ...will show PV OFS Use the up or down keys to enter an offset of 1 0 C Use the Home key to return to the original display You will now notice that the 1 0 C calibration has been added to the incubator t...

Страница 15: ...warranty service is required Thermoline Scientific will endeavour to have an authorised service organisation carry out the work if there is an available agent however if it is decided to return the e...

Страница 16: ...ly by Any defect in material or workmanship of or any other defect whatsoever in or unsuitability for any purpose of the goods or any part of the goods or default or negligence on the part of Thermoli...

Отзывы: