background image

Models TGO-80W, 150W, 240W             Glassware Drying Oven Manual 

26-1-16 

AL-2 

3.0ºC 

AT 

OFF 

AT.BS 

0.0 

 

 

 

ARW 

100.0 

CT 

BS 

 

LOCK 

 

  “BS” is used to adjust the PV value to make it read the true value when measured 

by a calibrated independent instrument. For example, if the “PV” display reads 
100.0ºC and a calibrated instrument measured the oven temperature at 100.5ºC; 

you would make “BS” = +0.5ºC using the up arrow. Don’t forget to press 

 to 

enter the value. 

 

Make the “LOCK” parameter 1 again 

before exiting the parameter configuration 

level by holding the 

 button for a few seconds

.

 

  The output of the temperature control will now adjust the oven to the correct 

calibrated temperature. 

WARRANTY: 

Have the following information available when you contact the service department: 

  Model number and serial number. This is generally found on the exterior of the case 

in the form of a stick-on label. 

 

 

  The company name, address, contact name, contact phone number. 

  A brief description of the problem. 

All warranty claims must be reported to, and agreed to by a Thermoline 
representative prior to any work being carried out. 
 

Содержание TGO-150W

Страница 1: ...Scientific Equipment Pty Ltd ABN 80 000 859 129 10 12 Ross Place Wetherill Park NSW 2164 P O Box 1851 Wetherill Park DC NSW 1851 Phone 02 9604 3911 International 61 2 9604 3911 Fax 02 9725 1706 Intern...

Страница 2: ...the packing material and check for damage Notify the detail of any damage to your supplier or to Thermoline Scientific without delay Retain the packing materials until the equipment has been thorough...

Страница 3: ...for the oven The TGO 80W and the TGO 150W are suitable for bench mounting but the larger TGO 240W is meant to be free standing Model TGO 80W TGO 150W TGO 240W Internal Dimensions 500 x 400 x 400 WDH...

Страница 4: ...he actual temperature in the oven SP set point This digital display is the set temperature in the oven This button is used to enter changes to parameters and to move between functions These arrows are...

Страница 5: ...ure will go through a short warm up before displaying the set temperature SP on the lower display and the oven temperature on the top display PV Use the arrow to highlight the digit on the set tempera...

Страница 6: ...ached the heating will turn off and the indicating light above the timer switch will be on Note to reset the timer you have to turn the main power switch off Press the button again to return to the or...

Страница 7: ...t accidental overheating Please be aware that the thermostat is a simple device that is less accurate than the digital controls It should not be set too close to the operating temperature in case it i...

Страница 8: ...the oven by a certified instrument For this reason the digital temperature control has a parameter that can be used to correct the temperature displayed called the BS In simple terms this BS adds or s...

Страница 9: ...ke the LOCK parameter 1 again before exiting the parameter configuration level by holding the button for a few seconds The output of the temperature control will now adjust the oven to the correct cal...

Страница 10: ...major failure Notwithstanding the preceding clause and to the extent permissible by law the liability of Thermoline is limited in relation to the warranted product and at the option of Thermoline to r...

Отзывы: