Models TGO-80W, 150W, 240W Glassware Drying Oven Manual
26-1-16
9
AL-2
3.0ºC
AT
OFF
AT.BS
0.0
P
I
D
ARW
100.0
CT
2
BS
LOCK
1
•
“BS” is used to adjust the PV value to make it read the true value when measured
by a calibrated independent instrument. For example, if the “PV” display reads
100.0ºC and a calibrated instrument measured the oven temperature at 100.5ºC;
you would make “BS” = +0.5ºC using the up arrow. Don’t forget to press
to
enter the value.
•
Make the “LOCK” parameter 1 again
before exiting the parameter configuration
level by holding the
button for a few seconds
.
•
The output of the temperature control will now adjust the oven to the correct
calibrated temperature.
WARRANTY:
Have the following information available when you contact the service department:
•
Model number and serial number. This is generally found on the exterior of the case
in the form of a stick-on label.
•
The company name, address, contact name, contact phone number.
•
A brief description of the problem.
All warranty claims must be reported to, and agreed to by a Thermoline
representative prior to any work being carried out.