SpeedVac™ SPD130DLX

 Operation  |   7

Audible Notification

The concentrator will play tones depending on the user 
actions. Below is a description of these tones. 

Manual Run

1. Connect the unit to its required voltage.

2. Turn the power switch located on the front of the UNIT, to 

the 

ON

 position, (light on switch indicates ON). The cover 

lock disengages, allowing the top cover to be opened.

The display lights up and shows the following in 
sequence:

Name of the concentrator

Software revision number

Default values:

-

Run Type

Time

-

Temperature

35°C

-

Run Time

2.00 hours

-

Vacuum 1 

Torr

3. Select a run configuration using one of the following 

methods:

Load a preset program

Load a user defined program

Directly modify values that are loaded on the user 
interface, using the 

Select

 button and the “

+

”/ “

-

” 

buttons.

4. Set the temperature set point between 35°C and 80°C, 

or “no”, for no heat.

5. Using the 

Select

 button and the “

+

”/ “

-

” buttons select 

and modify “Heat Time” to between 0.01 and 9.59 hours 
or “CCC” (for continuous heating). When the heat timer 
expires, the heater will shut off, no matter what the 
temperature set point reads (except if “CCC”).

6. Select Run Time: Since this is a manual run, no time 

adjustment is needed.

7. Select a vacuum set point: Since this is a manual run, the 

vacuum will be the maximum allowed by the attached 
external vacuum pump, regardless of the vacuum 
pressure set point.

8. Select the Run Type: Since this is a manual run, the run 

type setting is ignored. The vacuum will be the maximum 
allowed by the attached external vacuum pump, 
regardless of the vacuum pressure set point.

9. Place sample tubes in rotor so load is balanced. Secure 

rotor with the supplied knob (hand tight). Close cover.

10. Pre-heat may be selected at this time, to apply a 45°C 

heat to the chamber.

11. Press the 

Manual Run

 button. The cover locks and the 

lid locker indicator is illuminated. The rotor starts 
spinning. The “Run Time” display counts up. The 
temperature rises to the set temperature. The “Heat 
Time” will count down if set point is not set to “CCC” and 
if the temperature set point is not “no” The vacuum will be 
applied to the chamber and the level begins falling.

Note: 

If the cover is not closed, the display will show “Lid” 

and the run will not start.

12. Press 

RL On/Off

 for radiant chamber heat. Press at any 

time to turn ON and OFF. (As long as there is time left in 
the heat timer and if the temperature set point is not set 
to “no”).

13. To end the manual run, press 

Stop

 button. The unit will 

sound three audible beeps. The display will show “End”, 
the valves will click, isolating the chamber from the 
vacuum pump and allowing air to bleed into the chamber.

14. After the rotor stops spinning, the cover lock disengages 

and the lid locked indicator is turned off. The display 
reverts to last set parameters.

15. Open the cover and remove samples.

GENERAL: During the run, display shows actual parameters. 
To check set parameters, press 

View

 button and 

Select

. The 

Table 2. Audible Notifications

Event

Sound Description

Start-up

5 beeps

Starting a run (manual 
or automatic)

1 single beep

Manually stopping a 
manual or automatic 
run

3 beeps

Automatically 
stopping an automatic 
run

1 beep that is repeated until the 
user pushes the 

Stop

 button or 

opens the cover of the 
concentrator

Program saved

3 beeps

User input error 

1 long beep that lasts for 1 
second

System error

1 long beep that lasts for 1 
second and that is repeated 
until the user pushes the 

Stop

 

button or opens the cover of the 
concentrator

Содержание SpeedVac SPD130DLX

Страница 1: ...Installation and Operation 80302133 Revision A July 2018 SpeedVac SPD130DLX Vacuum Concentrator...

Страница 2: ...mo Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any p...

Страница 3: ...4 Installation of SPD130DLX 4 Operation 5 Description Of Control Panel 5 Manual Run 7 Auto Run Time 8 Auto Run Vacuum 8 Drying Rate 8 Preset Program Settings 9 Exporting Live Run Data 9 HyperTerminal...

Страница 4: ...th dual timer for automatic or manual control of run conditions it also features preset and user defined programs which enable user to retrieve and store application protocols for future use The compa...

Страница 5: ...dicates extreme low temperatures and high risk of frostbite Do not touch bare metal or samples with unprotected body parts This symbol indicates possible pinch points which may cause personal injury T...

Страница 6: ...tive Power 115 VAC 60 Hz 6 amps 230 VAC 50 Hz 3 amps Vacuum Chamber TEFLON coated chemical resistant aluminum casting Cover Heated cover with radiant lamps includes a safety interlock Induction Motor...

Страница 7: ...130DLX units configured for 230 VAC 50 Hz should be plugged into a circuit rated for at least 3 amps Note Do not use any detachable power cord that is not adequately rated for the unit Installation of...

Страница 8: ...hand to ensure alignment of the pin with the groove Secure the assembly by screwing the retaining knob into the drive shaft above the rotor Tighten it firmly but not excessively Description Of Contro...

Страница 9: ...meters is shown by the corresponding indicator that is illuminated 8 Minus Decrements the value of the selected parameter 9 Plus Increments the value of the selected parameter 10 View During a run pre...

Страница 10: ...tor with the supplied knob hand tight Close cover 10 Pre heat may be selected at this time to apply a 45 C heat to the chamber 11 Press the Manual Run button The cover locks and the lid locker indicat...

Страница 11: ...and remove samples Auto Run Vacuum 1 Refer to the Manual Run section for start up 2 To execute an Automatic Vacuum run execute the following steps a Select the Run Type as Vac b Use the Select button...

Страница 12: ...C The heat time is expressed as a 3 digit integer in minutes or CCC for continuous heating The runtime is expressed as a 3 digit integer in minutes The vacuum pressure is represented as a floating poi...

Страница 13: ...id stay mechanism that will assist the user in opening and closing the cover of the unit It will secure the lid and maintain it open allowing the user to operate the unit with both hands Figure 3 Lid...

Страница 14: ...cid Ammonium hydroxide n butyl alcohol Chloroform Cyclohexane DCM Ethyl acetate Ethanol Ethyl ether n Hexane n Heptane Isopropyl alcohol Isobutyl alcohol Methanol Petroleum ether n propyl alcohol Pyri...

Страница 15: ...0mm RH32 13 15 8 17 5 X 102 mm 15ml Corex tubes RH8 17 5 10 8 18 X 100 mm 17 x 95 16 x 100 RH8 18 CENTRIFUGE TUBES 12 10 15ml conical 16 x 120 mm RH10 15 40 6 50ml conical 28 x 115 mm RH6 50 FLASKS 80...

Страница 16: ...neutralizing solution for Oligo Preps Used on UVS850DDA only SCT120 Chemical trap order cartridges separately DTK120R Chemical trap kit for radioactivity compatible with all Thermo vacuum concentrator...

Страница 17: ...Pump ANT100 Exhaust recovery Post trap Vacuum Tubing From Cold Trap to Back of SPD130DLX UVS850DDA Vacuum Tubing From Back of SPD130DLX To Cold Trap SpeedVac Concentrator SPD130DLX Vacuum Tubing RVT45...

Страница 18: ...MA number issued by Thermo Costs of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardo...

Страница 19: ...riciclato tramite queste medesime Ulteriori informazioni sulla conformit di Thermo Fisher Scientific con queste Direttive l elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodo...

Страница 20: ...30 France 33 2 2803 2180 33 2 2803 2180 Germany 0800 1 536 376 49 6184 90 6000 0800 1 536 376 Italy 32 02 95059 552 39 02 95059 552 432 254 375 Netherlands 31 76 579 55 55 31 76 571 4440 Nordic Baltic...

Отзывы: