SpeedVac™ SPD130DLX
Operation | 5
Operation
Connect the power cord to the power inlet on the back of the
unit and plug it into the appropriate outlet. Turn on the main
power switch located on the front of the unit in the lower right
to energize the unit. The safety lid locking mechanism on the
SPD130DLX will automatically disengage. Do not attempt to
open the lid unless the unit is plugged in.
Note:
The SPD130DLX Concentrator is an important
component in a complete concentration system that should
also include a refrigerated vapor trap and a vacuum pump
along with other applicable accessories. Please refer to
operating instructions of each component for details on
usage.
Rotor installation.
Open the lid of the rotor chamber.
Visually align the pin on the drive shaft with the groove on the
bottom of the rotor. Carefully lower the rotor onto the drive
shaft. Rotate the rotor by hand to ensure alignment of the pin
with the groove. Secure the assembly by screwing the
retaining knob into the drive shaft above the rotor. Tighten it
firmly but not excessively.
Description Of Control Panel
1.
Lid locked
- Indicates whether or not the lid of the
concentrator is locked. The indicator will be illuminated
when locked.
2.
Preset buttons
– 3 preset programs that are not
modifiable.
Loading a preset is achieved by pressing the
corresponding
Preset
button. The following parameters
are automatically loaded into the user interface:
•
Temperature set point
•
Heat time
•
Run time
•
Vacuum pressure
•
Run type
CAUTION:
Use appropriate Thermo Fisher
Scientific designed rotors recommended for the
Product.
CAUTION:
Load the rotor and close the lid.
Always balance rotor loads. An imbalanced rotor
causes vibration that will damage the system’s
bearings and equipment. Load the rotor
symmetrically. There need not be a tube in each
holder, but the load must be evenly distributed.
When using a rotor with aluminum tube holders,
insert all the tube holders.
1
4
5
6
11
12
15
16
14
13
10
9
8
7
2
3