21
Running E-PAGE
™
Gels,
Continued
Using E-Base
™
,
Continued
3.
The Mother E-Base
™
Device
or Daughter E-Base
™
Device
signals the end of
the run with a flashing red light and rapid beeping for 2 minutes followed
by a single beep every minute.
At the end of the run, the digital display shows the original time setting.
The digital display does not any time change that was made during
electrophoresis. The digital display also shows the elapsed time (up to
19 minutes with a negative sign) since the end of the run.
4.
Press and release
the pwr/prg button to stop the beeping. The light turns
to a steady red and the digital display shows the last time setting.
5.
Remove the gel cassette from the Mother E-Base
™
Device
or Daughter
E-Base
™
Device.
6.
For visualizing Lumio
™
fusion proteins, see page 38. For opening the
E-PAGE
™
cassette, see page 24.
Note: The bands in the gel will diffuse within 40 minutes.
Interrupting an
Electrophoresis
Run
You can interrupt an electrophoresis run at any time by
pressing and releasing
the pwr/prg button to stop the current. The stopped current is indicated by a
steady red light and the digital display will flash to indicate that the run was
interrupted.
You can remove the gel from the E-Base
™
Device to check the progress of the run.
Then:
•
To
continue
the run from the point at which it was stopped, reinsert the gel
and press and release the pwr/prg button. The light changes to steady green
and the digital display shows the count down time.
•
To
cancel
the rest of the interrupted run, press and hold the pwr/prg button
for a few seconds. The digital display will reset and the base will return to
Ready Mode. If desired, you can then program a new run time as described
on page 15 and rerun the gel.
In case of an
external
power failure
(loss of electricity or the electrical cord is
accidentally removed from the outlet), the run will continue when the power
resumes. The Mother E-Base
™
Device
or Daughter E-Base
™
Device
signals the end
of the run as described on the previous page, except the light will be an
alternating red/green to indicate that an external power failure had occurred
during the run.
We recommend that you disconnect the Mother E-Base
™
Device from the
electrical outlet when not in use for a prolonged period of time.
Maintaining
E-Base
™
Keep the surfaces of the Mother E-Base
™
Device and Daughter E-Base
™
Device
free of contaminants. To clean, disconnect bases from power source and wipe
with a dry cloth. Do not attempt to open or service the bases. To honor the
warranty, bases should only be opened and serviced by Life Technologies.