15
Electrophoresis of E-PAGE
™
Gels,
Continued
Selecting a
Program on the
E-Base
™
Device
The recommended program for E-PAGE
™
Gels is EP.
Select program EP prior
to inserting a gel into the base.
Brief instructions for using the E-Base
™
Device are included in this manual. For
details, refer to the E-Base
™
manual.
1.
Plug Mother E-Base
™
Device into an electrical outlet using the electrical
plug on the base. Connect the Daughter E-Base
™
Device to a Mother E-
Base
™
Device or another Daughter E-Base
™
Device connected to a Mother
E-Base
™
Device.
2.
The display shows EP or last program used (EG or EP).
3.
Press and release the pwr/prg (power/program) button to select the
program EP
. The default run time for program EP is 14 minutes.
4.
To manually change the default run time, press and release the time button
located on the lower-right corner of the base to view the time setting.
5.
Press and hold the time button to increase the time to the desired run time.
If the time button is not released, the time setting increases until it reaches
00. To begin cycling through the numbers again, starting from 00, press the
time button again.
Gel Type
Time in Minutes
E-PAGE
™
48
8%
25 minutes
E-PAGE
™
96
6%
14 minutes (default run time for program EP)
Do not run an E-PAGE
96
Gel for more than 25 minutes, or an E-PAGE
48
Gel for more than 30 minutes.
•
When performing electrophoresis of Lumio
™
fusion proteins, we
recommend extending the run time of the gel for an additional 2 minutes
to prevent the formation of a fluorescent dye front.
•
If your sample contains high salt or detergent concentrations, you may
need to manually increase the run time.
•
To increase the run time when a cassette is inserted, press and release the
time button to increase the time setting by 1-minute intervals or press and
hold the time button to increase the time continuously.
•
To increase the run time while a run is in progress, see next page. To
manually interrupt or stop a run, see page 21.