26
Using the iBlot
®
Device with the De-Bubbling Roller,
Continued
Assembling the
iBlot
®
Device
1.
Open the lid of the device and pull up the Metal Spacers 1 and 2. If you
have attached the De-bubbling Roller to the device, then remove the roller
as shown in the figure below.
2.
Remove the package labeled iBlot
®
Anode Stack, Bottom from the iBlot
®
Gel
Transfer Stacks Box. Remove the laminated sealing of the iBlot
®
Anode
Stack, Bottom and
keep the stack in the transparent plastic tray.
3.
Place the iBlot
®
Anode Stack, Bottom stack
with the tray
to the left of the
blotting surface area such that the tab on the tray is on the right side of the
De-bubbling Roller, as shown below. Slide the bottom stack to the left until
the stack is blocked by the Gel Barriers present on the left side of the device.
Note:
Handle the iBlot
®
Anode Stack, Bottom using the plastic tray to avoid
disturbing the gel and membrane layers in the stack.
Do not
touch the
transfer membrane on the stack.
Spacer 2
Roller
Tab