3.
Select a temperature field, then enter the desired temperature. Temperatures for neighboring
VeriFlex
™
Blocks can be the same, but if they are not the same, they must be
at least
0.1°C
apart but
no more
than 5.0°C apart. The maximum temperature is 99.9°C.
4.
for each of the six VeriFlex
™
Blocks, then select
Done.
Note:
To set each block to the same temperature, select the desired temperature field, then select
Apply Selected to All
.
5.
Depending on your instrument configuration, enter either a duration or a ramp rate.
• Select
Time
, enter the duration (in minutes and seconds), then select
Done
.
or
• Select
Ramp Rate
, enter the ramp rate (as a percentage value), then select
Done
.
The VeriFlex
™
Step appears in the
Edit Run Method
screen, with the text “Zones” in the
temperature field.
Chapter 3
Performing a run
Editing details
3
56
Veriti
™
Thermal Cycler User Guide