Thermo Scientific
Chapter 6
|
Specifications and part numbers
HyPerforma Smartainer 3.0 User’s Guide
|
50
Table 6.2. 500 L Smartainer 3.0 unit specifications.
Specification
Jacketed
Non-jacketed
Un
it ge
om
et
ry
Rated liquid working volume
500 L
Total volume (liquid and gas)
550 L
Interior length
114.3 cm (45 in.)
Interior width
73.66 cm (29 in.)
Interior height
88.64 cm (34.9 in.)
Overall length
125.22 cm (49.3 in.)
124.46 cm (49 in.)
Overall width
84.07 cm (33.1 in.)
83.82 cm (33 in.)
Overall height
125.73 cm (49.5 in.)
Extra height (for lid)
81.02 cm (31.9 in.)
Te
mp
er
at
ur
e c
on
tr
ol
Jacket area
19,994.79 cm
2
(3,099.2 in
2
)
N/A
Jacket volume
9.46 L
N/A
Maximum jacket pressure
5.17 bar (75 psig)
Jacket flow rate at 3.4 bar (50 psi)
109.5 L/min.
N/A
Process connection
1.5 in. Sanitary tri-clamp
N/A
Recommended TCU model: maximum heating/cooling
TF10000: 6100/10000 W
N/A
Approximate liquid heat-up time, hours:minutes
(5°C–37°C)—100% volume
3:46 (3:43 with circulation)
N/A
Approximate liquid cool down time, hours:minutes
(37°C–5°C)—100% volume
10:41 (10:47 with circulation)
N/A
Gen
er
al
Dry skid weight* (mass)
383 kg (845 lbs)
304 kg (670 lbs)
Wet skid weight*—rated working volume (mass)
883 kg (1,947 lbs)
804 kg (1,772 lbs)
Pneumatic supply requirement, where applicable
Minimum 5 bar (72.5 psi)
Required tubing size for hoist pneumatic air line
8 mm
Load cell display power supply requirement (voltage,
phase, amp)
85–264 VAC, single, 10/20 A
Temperature display power supply requirement (voltage,
phase, amp)
85–264 VAC, single, 10/20 A
Noise level
< 70 dB at 1.5 m
Re
comm
en
de
d
op
er
at
in
g
par
am
et
er
s
Environmental operating temperature range
Ambient to 40°C ± 0.5°C (104°F ± 0.9°F)
Volume range
0–500 L
Maximum BPC pressure
0.03 bar (0.5 psi)
*Vessel-only estimate. Accessories will add additional weight.