Thermo Scientific
D-4 ThermoFlex
Appendix D
Thermo Scientific
Thermo Scientific
Analog Values
Qualifier Byte
b.7
b.6
b.5
b.4
b.3
b.2
b.1
b.0
A qualifier byte of 0x12 indicated that the value contains one decimal point and the units
are °F , e.g., 98.6°F.
Example to set setpoint to 25°C:
If the temperature units are unknown, before changing the setpoint send a command to request
setpoint. The response will include both the precision and units. Precision is fixed at 0.1 and units can
be either °C or °F. If the units are already known skip to step 3.
1. Master sends:
CA 00 01 70 00 8E
(REQ SETPOINT1)
2. Slave responds:
CA 00 01 70 03 11 00 C8 B2 Precision =0.1, units =°C, value=200
(200 x 0.1°C=20.0°C)
Response indicates:
uses a 2 byte integer
(nn=03)
precision and units are 0.1°C
(d1=11)
3. Master sends:
CA 00 01 F0 02 00 FA 12
(Set Setpoint 1 to 25.0°C)
4. Slave responds:
CA 00 01 F0 03 11 00 FA 00 Precision =0.1, units =°C, value=250
(250 x 0.1°=25.0.0°C)
See Additional Command Examples in this Appendix.
Unit of Measure
Index
Unit
0
NONE
1
Temperature in °C
2
Temperature in °F
3
Flow liters per minute
4
Flow in gallons per minute
5
Time in seconds
6
Pressure in PSI
7
Pressure in bars
8
Resistivity in M
Ω
-cm
9
%
10
Volts
11
Pressure in kPa
Precision of measurement
Unit of measure index
Содержание ThermoFlex10000
Страница 2: ......
Страница 36: ...ThermoFlex900 5000 Thermo Scientific...
Страница 42: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Страница 79: ...517 2014 16 2014 ThemoFlex24000...
Страница 100: ...Section 2 2 20 ThermoFlex Thermo Scientific...
Страница 136: ...Section 4 4 14 ThermoFlex Thermo Scientific...
Страница 168: ...Section 7 7 14 ThermoFlex Thermo Scientific...
Страница 191: ...C 10 ThermoFlex Appendix C Thermo Scientific...
Страница 208: ......
Страница 209: ......
Страница 210: ...Thermo Fisher Scientific 168 Third Avenue Waltham Massachusetts 02451 United States www thermofisher com...