20
error Disable - err
This feature is available for all units, but not all sub-menu
options will be applicable since some errors relate to heating
or stirring only. The purpose of this feature is to enable or
disable a particular error from being detected and displayed.
all errors except e12 (locked rotor – stirring control) are able
to be disabled. a general option is also available to re-enable
all errors at once. When disabling errors though they must be
done one at a time.
To modify/view the error Disable menu follow the instructions
given below:
1.
plug the unit into the appropriate power, but do not
turn it on (display should be blank), enter the
Service Menu by pressing and holding the poWer
key. after about 3 seconds a single beep will be
heard and you may remove your finger from the
poWer key. The display will change to
oSP
.
2.
rotate the center knob until the display reads
err
.
press the SeT key under the display to accept, and
the display will change to
Clr
(re-enable all disabled
errors). rotate the center knob to change the dis-
play to the error number of choice and press the
SeT key under the display again.
3.
The display will now indicate either
on
(the error
detection is enabled), or
off
(the error detection is
disabled). rotate the center knob to change the dis-
play if desired and press the SeT key under the dis-
play.
4.
The unit will return to the previous menu. Select
another error to change, or press the poWer key
to return to the root Service Menu. pressing the
poWer key one more time will turn the unit off.
S
erVICe aND
C
alIBraTIoN
Содержание Super-Nuova S133320-33
Страница 24: ...24 exploded views 10x10 Stirrer Exploded View 220 240v models only 1 2 3 4 5 6 7 8 15 11 14 13 12 16 9 10...
Страница 25: ...25 eXPloDeD VIeWS 10x10 Stirrer Top Assembly 17...
Страница 26: ...26 eXPloDeD VIeWS 6 5 7 8 9 10 11 12 13 14 16 15 1 2 3 4 7x7 Stirrer Exploded View 220 240v models only...
Страница 27: ...27 eXPloDeD VIeWS 17 7x7 Stirrer Top Assembly...
Страница 32: ...32...