19
Timer Shutdown - end
This feature is available for all units, but not all sub-menu
options will be applicable. The purpose of this feature is
to control what is turned off when the timer elapses. The
default is heating off for hot plates and stir plates, and
stirring off for stirrers.
To modify/view the Timer Shutdown follow the instructions
given below:
1.
plug the unit into the appropriate power, but do
not turn it on (display should be blank), enter the
Service Menu by pressing and holding the
poWer key. after about 3 seconds a single
beep will be heard and you may remove your
finger from the poWer key. The display will
change to
oSP
.
2.
rotate the center knob until the display reads
end
. press the SeT key under the display to
accept, and the display will change to H
(Heating off), H S (Heating and Stirring off), or S
(Stirring off). rotate the center knob to change
the display to the method of choice and press
the SeT key under either display again.
3.
The unit will return to the previous menu. Select
another feature to change, or press the poWer
key again to return to the off mode.
S
erVICe aND
C
alIBraTIoN
Note
For most routines, the poWer key
can be used to return to a previous
menu if desired. also, many of the
options can be checked to see how
the unit is configured by selecting
the feature and observing which
menu item is displayed. The unit
will always display the current con-
figuration first.
Содержание Super-Nuova S133320-33
Страница 24: ...24 exploded views 10x10 Stirrer Exploded View 220 240v models only 1 2 3 4 5 6 7 8 15 11 14 13 12 16 9 10...
Страница 25: ...25 eXPloDeD VIeWS 10x10 Stirrer Top Assembly 17...
Страница 26: ...26 eXPloDeD VIeWS 6 5 7 8 9 10 11 12 13 14 16 15 1 2 3 4 7x7 Stirrer Exploded View 220 240v models only...
Страница 27: ...27 eXPloDeD VIeWS 17 7x7 Stirrer Top Assembly...
Страница 32: ...32...