The technical data of the Sorvall Legend X1R is listed in the following table.
Table 1-1.
Technical Data Sorvall Legend X1R
Feature
Sorvall Legend X1R
Environmental conditions
-Use in interior spaces
-Altitudes of up to 2,000m above sea level
-max. relative humidity 80% up to 31 °C; decreasing linearly up to 50%
relative humidity at 40 °C.
Permissible ambient temperature
+2 °C to +
3
5 °C
Overvoltage category
II
Pollution degree
2
Voltage
230 V
Heat dissipation
4778 BTU/h
IP
20
Run time
unlimited
Max speed n
max
15200 rpm (depending on the rotor)
Min speed n
min
300 rpm
Maximum RCF value at n
max
25830 x
g
Maximum kinetic energy
62.5 kJ
Noise level at maximum speed
< 63 dB (A)
Temperature setting range
-10 °C to +40 °C
Dimensions
refrigerated
Height
Height with lid open
Width
Depth
Table top height
360 mm
870 mm
623 mm
605 mm
310 mm
Weight without rotor
91.5 kg
Order No.
Refrigerant
Quantity Pressure
GWP CO2e
75004260
R-404A
0.38 kg
31 bar
3922 1.49 t
Centrifuge
Thermo Scientific
Sorvall Legend X1R
Contains fluorinated greenhouse gases in a hermetically sealed system.