4
Operation
Centrifugation
4-10
Sorvall Legend X1R
Thermo Scientific
The display restores the current values. Return to step 1.
Centrifugation
Once the rotor has been properly installed, the main switch has been turned on and the lid has been
closed, you can start centrifuging.
Starting centrifuge program
Press the
key on the control panel. The centrifuge accelerates to the preset speed with the time
display active.
If the speed setting is higher than the maximum permissible speed or RCF-value for the particular
rotor, then the display will show the message
CHECK SET SPEED
. The centrifuge decelerates. Please
enter a new value.
You cannot open the lid as long as the centrifuge is running.
Imbalance indicator
If a load is imbalanced, this will be indicated at speed higher than approx. 300 rpm by the message
Imbalance load
.
The run will terminate.
Check the loading and start the centrifuge once again. See the information on proper loading in the
rotor instruction manual. For information on troubleshooting, see section
“Troubleshooting by user”
on
page 7-3
.
Stopping the centrifugation program
With preset run time
Usually the run time is preset and you only have to wait until the centrifuge stops automatically when
the preset time limit expires.
As soon as the speed drops to zero, the message
End
will appear in the display. By pressing the
key,
you can open the door and remove the centrifuge material.
You can also stop the centrifuging program manually at any time by pressing the
key.