1
Introduction and description
Functions and features
Thermo Scientific
Sorvall Legend X1R
1-5
Functions and features
Mains supply
The following table contains an overview of the electrical connection data for the Sorvall Legend
X1R. This data is to be taken into consideration when selecting the mains connection socket.
The following table gives an overview of the important functional and performance characteristics
of the Sorvall Legend X1R.
Table 1-3.
Functions and Features
Component / function
Description / features
Structure / Housing
Galvanized steel chassis with armoured plating
Chamber
Stainless steel
Drive
Induction drive without carbon brushes
Keys and display
Easy-to-clean keypad and display surface
Controls
Microprocessor-controlled
Internal memory
The most recent data is saved
Functions
RCF-selection, temperature control, pretemp with cooling equipped
devices
Acceleration / braking profiles
9 acceleration and 10 braking curves
Rotor recognition
Automatic
Imbalance recognition
Electronic, contingent on rotor and speed
Lid lock
Automatic lid closing and locking starting from an initial hold position
Table 1-4.
Electrical connection data of the Sorvall Legend X1R
Cat.
Mains
voltage
Frequency
Rated current
Power
consumption
Equipment
fuse
Building fuse
1400 W
15 AT
16 AT
refrigerated
230 V
±10 %
50 / 60 Hz
8 A
1
Please contact the Thermo Fisher service to run the unit at 60Hz L
ine frequency.