4-2
4.3.1 Alarms List
Clock Fail
The system has detected a failure on the clock calendar circuit. (Only with COM Board
option)
Go to the
DIAGNOSTICS
screen and re-enter the date and time.
Check the battery
Replace the motherboard.
Load Cell Fail S#
S# identifies the scale if more scale are defined. The system has detected an error on the
load cell signal.
Check the load cell connections.
Check the load cell(s).
RAM Fail
The system has detected an error on the
RAM
(
Random Access Memory
) checksum during
the internal periodic test. The
RAM
is used to store variables and set up data.
Replace the motherboard.
ROM Fail
The system has detected a failure on the
ROM
(
Read Only Memory
)
checksum
during the
internal periodic test. The
ROM
is used to store the program.
Replace the mother board
Threshold #1 S#
S# identifies the scale if more than one scale is defined. The threshold 1 has been reached.
Threshold #2 S#
S# identifies the scale if more than one scale is defined. The threshold 2 has been reached.
Threshold #3 S#
S# identifies the scale if more than one scale is defined. The threshold 3 has been reached.
Threshold #4 S#
S# identifies the scale if more than one scale is defined. The threshold 4 has been reached.
Warm Start
The system has detected a power loss condition, or power was removed for an undefined
period.
Cold Start
The system has detected the loss of the set up data after power was removed. The
instrument needs to be setup and calibrated.
Replace the motherboard
Содержание Ramsey MINI CK100
Страница 1: ...1 Operating And Service Manual Ramsey MINI CK100 Weight Indicator MINI CK100 Part No...
Страница 4: ......
Страница 14: ......
Страница 24: ......
Страница 80: ......
Страница 86: ......
Страница 92: ......
Страница 94: ......
Страница 95: ...C 1 Appendix C Drawings Wiring Diagram CK 0000100 E001C...
Страница 96: ......
Страница 97: ......