3.8 Calibration
The
MAIN
MENU 1
contain the Calibration Menu. The MENU 1 is selected by pressing
SETUP. Desired calibration scrolls are selected by pressing the soft keys direcly below the
desired scroll.
3.8.1 Zero Calibration Scroll
The Zero calibration is implemented as a machine directed procedure
1.
Press
ZERO CALIB
, the next screen is shown:
During Auto Zero procedure the scale must be kept empty. A complete zeroing procedure
requires 10 seconds, but can be reduced by pressing END in the next scroll.
Press
MANUAL
for next functio 2
Pressing
RUN
returns to menu
RUN
2.
Press
START,
the next screen is shown
:
During
Auto Zero
, weight resolution is 10 times higher than normal. The number of
seconds in line 2 corresponds to the time remaining for completing the test.
-- MENU 1 --
Premi MENU for more
ZERO SPAN LINEAR
CAL CAL
-- ZERO CAL --
Empty scale, then press
START
START MANUAL
AUTO ZEROING
Time remaining 0000
Gross: 000.0 kg
END ABORT