The system calculates the CALCON (Calibration Constant) based on the mechanical and
electrical parameters entered in the Scale Data Scroll.
This menu is for reference only.
The R-Cal factor can be computed during the autospan function and used to correct the error
between the two span methods.
Default:
INVALID
(0)
Min:
-99,99%
Max:
+99,99%
3.11.2
Calibration Interval (Option COMM)
The system can be programmed to prompt the operator when a certain amount of time has
passed since the last calibration. If you do not want to use this option, confirm the default 0
days interval, otherwise enter the number of days. The calibration date displayed in Scroll 9 is
automatically updated whenever a calibration is performed. If a non zero value is entered, an
alarm appears after the time is elapsed. The alarm can only be cleared after a calibration check
is executed.
Default:
0
Days
Min:
0
Days
Max:
365
Days
-- CAL DATA SCROLL 4 --
R-CAL Constant
00.00
MENU
-- CAL DATA SCROLL 5 --
RCAL Factor
Invalid
ENTER +/-
Password:Service
--CAL DATA SCROLL 6 --
Calibration Interval
365 Days
ENTER
Password:Operator