RadEye GX, GX-L
DB-074 E
15-3
Pi/Ff 14.09.2016
Power consump-
tion:
1.4 mA: normal operation without alarm
signals and LCD illumination. High voltage:
900V
27 mA with illuminated LC display
9 mA LED alarm
33 mA acoustic alarm
26 mA vibrator alarm
Battery service
life:
500h using two alkaline AAA cells de-
pending on the operating mode
250h using 800mAh NiMH accumulator
15.2
RadEye GX-L
Measured
quantities:
Count rate (cps, cpm)
Surface contamination (Bq, dps, dpm, Bq/cm
2
)
Dose rate (R/h, Sv/h, rem/h, Gy/h)
Measuring range:
Max. 10,000 cps
Largest indication
Count rate: 10 kcps, 600 kcpm,
Activity: 9999 kBq, 9999 kdps, 9999 kdpm,
9999
kBq/cm²
Dose rate: 9999 R/h, 9999 rem/h, 9999 Sv/h
Probe cables
RG 58 or RG62, Max. 2 m – MHV connector,
High voltage range
350V…1100V
Output impedance
1M
Ω