RadEye GX, GX-L
DB-074 E
3-13
Pi/Ff 14.09.2016
Pressing the
/dose – key shows upon the
- 1
st
click:
the accumulated dose (rem, R or Sv) is shown
only if dose rate mode (rem/h, R/h, or Sv/h) is ac-
tive. It displays the accumulated dose and the re-
maining time until the dose alarm 1 is reached,
assuming the current dose rate will persist.
- 2
nd
click:
the mean value and max value (“Peak”) of the
measurement together with the elapsed time indi-
cated since the last reset (hhh:mm:ss).
- 3
rd
click:
standard display
After 10 seconds or after 3
rd
click standard display with normal
rate indication is activated again.
1
st
click (only in dose rate mode)
2
nd
click
The RadEye supports two Ratemeter modes:
Ratemeter ADF
: Ratemeter with filter type:
A
dvanced
D
igi-
tal
F
ilter (ADF). This is a digital RC-Filter with time con-
stant from 1s to 180s, depending on count rate and count rate
changing.
Ratemeter Tau
: Ratemeter with linear filter. The time con-
stant is settable via menu item “Edit Tau” (see chapter 4.17)
from 1s to 60s
A long key press on the menu button resets the Ratemeter to the
actual count rate.