Troubleshooting Guide
Warning and error codes
54
Thermo Scientific PikoReal Real-Time PCR System User Manual
Thermo Fisher Scientific
Code
Symbolic name
Description
Action text
E33
SELF TEST FAILURE
This error is set if the RT self test fails.
The self test is run at every boot and
protocol start and it checks that the
camera is found, the filter is initialized
and the plate is configured (for a final
tested instrument).
REBOOT
INSTRUMENT. IF
PROBLEM PERSISTS
CONTACT SUPPORT
E35
LID TEMP. CHANGE
This error is set if the lid is active, the
temperature difference to the target is
more than 20°C and the response to a
10 seconds heating is less than 2°C.
CONTACT SUPPORT
E36
STRAIN GAUGE FAILURE
This error is set if the lid pressure
sensor is broken.
CONTACT SUPPORT
E37
LID OPTO FAILURE
This error is set if the lid position
sensor is unable to detect the home
position.
RETRY. IF PROBLEM
PERSISTS SEE
MANUAL
E40
DRAWER FAILURE
This error is set if the drawer is
opened, but the drawer opto state is
not changed in 100 ms.
CONTACT SUPPORT
Cont.
Содержание PikoReal
Страница 1: ...Thermo Scientific PikoReal Real Time PCR System User Manual Rev 2 0...
Страница 2: ......
Страница 3: ...Thermo Scientific PikoReal Real Time PCR System User Manual Rev 2 0 Cat no N11471...
Страница 12: ...Contents 10 Thermo Scientific PikoReal Real Time PCR System User Manual Thermo Fisher Scientific...
Страница 62: ...System Log 60 Thermo Scientific PikoReal Real Time PCR System User Manual Thermo Fisher Scientific...
Страница 66: ...Glossary 64 Thermo Scientific PikoReal Real Time PCR System User Manual Thermo Fisher Scientific...
Страница 70: ...68 Thermo Scientific PikoReal Real Time PCR System User Manual Thermo Fisher Scientific Notes...
Страница 71: ......
Страница 72: ...Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com N11471...