24
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
LCD
Message
Description
Corrective Actions
ERR 8
Excessive stray light detected.
Verify that the vial is fully in the
sample well and light shield cover
is properly placed over the vial.
Attempting a measurement when
low battery indicator is on.
Replace the batteries.
ERR 9
Lamp failure.
Contact Technical Support.
Or
Turbidity value is above the
measurement range (>1000 NTU).
Dilute sample, see the
Dilution
section.
Low battery indication – the
batteries need to be replaced.
Replace the batteries.
Note:
If an error message appears, take the
appropriate corrective action and re-do the
desired procedure . If the problem persists,
contact Technical Support or your dealer .
After troubleshooting all components of your measurement system,
contact Technical Support . Within the United States call 1 .800 .225 .1480
and outside the United States call 978 .232 .6000 or fax 978 .232 .6031 .
In Europe, the Middle East and Africa, contact your local authorized
dealer . For the most current contact information, visit
www .thermo .com/contactwater .
For the most current warranty information, visit www .thermo .com/water .
The AQ3010 meter is supplied with a 3 year warranty from
manufacturing defects and calibration standards for 6 months . If
repair or adjustment is necessary and has not been the result of abuse or
misuse within the designated period, please contact Technical Support
or your dealer for an RMT number . A repair technician will determine
if the product problem is due to deviations or customer misuse . Out of
warranty products will be repaired on a charged basis .
The instrument warranty shall not apply to defects resulting from:
• Improper or inadequate maintenance by customer
• Unauthorized modification or misuse
• Operation outside of the environment specifications of the products
Waterproof Seal:
Opening the instrument enclosure (excluding the
battery compartment) may void the warranty .
Assistance
Warranty
Exclusions