Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
23
Chapter 5
Customer Services
The AQ3010 turbidity meter routinely performs self-diagnostics and will
automatically generate messages to provide you with specific diagnostic
information . These messages are for your use and do not indicate a
reduction in the performance of the instrument or a failure of any
component in the instrument, unless otherwise stated in this list .
LCD
Message
Description
Corrective Actions
ERR 1
Calibration error – the meter is
unable to recognize the 800 NTU
calibration standard.
Use the correct 800 NTU
calibration standard. Pour fresh
standard into a clean vial and retry
the calibration. Verify that the
standard has not expired.
ERR 2
Calibration error – the meter is
unable to recognize the 100 NTU
calibration standard.
Use the correct 100 NTU
calibration standard. Pour fresh
standard into a clean vial and retry
the calibration. Verify that the
standard has not expired.
ERR 3
Calibration error – the meter is
unable to recognize the 20.0 NTU
calibration standard.
Use the correct 20.0 NTU
calibration standard. Pour fresh
standard into a clean vial and retry
the calibration. Verify that the
standard has not expired.
ERR 4
Calibration error – the meter is
unable to recognize the 0.02 NTU
calibration standard.
Use the correct 0.02 NTU
calibration standard. Pour fresh
standard into a clean vial and retry
the calibration. Verify that the
standard has not expired.
ERR 5
Calibration error – there is not
sufficient signal to achieve the
appropriate resolution in the 0 to
1000 NTU range.
Pour fresh standard into a clean
vial and retry the calibration with
all four standards. Verify that the
standards have not expired.
ERR 6
General calibration failure – there
is not sufficient signal to achieve
the appropriate resolution in the 0
to 100 NTU range.
Pour fresh standard into a clean
vial and retry the calibration with
all four standards. Verify that the
standards have not expired.
ERR 7
General calibration failure – there
is not sufficient signal to achieve
the appropriate resolution in the 0
to 20 NTU range.
Pour fresh standard into a clean
vial and retry the calibration with
all four standards. Verify that the
standards have not expired.
Troubleshooting
Guide