6
External Devices
Communication Connectors
18
Orbitrap Fusion Series Getting Connected Guide
Thermo Scientific
To connect the contact closure cable
1. Connect the contact closure cable to the
and Ground pins on the left side of the
2. Connect the other end of the cable to the external device according to its manual.
Communication Connectors
Use the left-side communication connectors to connect the MS to the Ethernet switch, the
syringe pump, the modular valve, and any external device.
lists the pin-out descriptions for the connectors shown in
.
IMPORTANT
• Before proceeding, verify that the external device is suitable for use with the
Orbitrap Fusion Series MS. Make sure that the signal from the external device meets
the requirement stated in the pin 4 (
pin-out description (see
). If it
does not meet this requirement, you cannot use it with the MS.
• To help prevent sample waste if the MS becomes unresponsive, connect a cable
between the optional LC device and the MS’s Ready Out and Injection Hold pins (see
). Then, configure the Contact Closure settings in the Method Editor. You are
responsible for providing this cable.
Table 3.
Pin-out descriptions for the communication connectors (Sheet 1 of 3)
Pin
Name
Description
–
Reset
Resets the instrument to a power-up state.
Note
Use this button only if the instrument does not
respond to the control program on the data system
computer or if you need to restart the instrument
without turning off the electronics service switch.
Peripheral Control
1
Ground
Earth ground.
2
5V
Provides a 5 Vdc, 500 mA output (with pin 1).
Содержание Orbitrap Fusion Lumos MS
Страница 1: ...Orbitrap Fusion Series Getting Connected Guide 80011 97002 Revision C June 2017...
Страница 6: ......
Страница 8: ...Contents viii Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...
Страница 24: ...4 Syringe Pump and Modular Valve Assembly 12 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...
Страница 28: ...5 Electrical Connections 16 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...