2
Gas Connections
6
Orbitrap Fusion Series Getting Connected Guide
Thermo Scientific
To connect the gas lines to the back of the MS
IMPORTANT
• After you start using the MS, do not shut off the gases. Optimum performance requires a continuous gas
flow.
• Do not shut off the UHP nitrogen gas unless you have vented the MS.
• If you intend to use helium for sparging your LC solvents, you must have a second tank and regulator.
Swagelok fittings, 1/8 in. ID
(UHP helium gas) Repeat Step 2 with another length
of copper tubing, except connect it between the
Helium In
inlet and your UHP helium gas supply.
3
(UHP nitrogen gas) Cut a length of the copper tubing,
attach the 1/8 in. ID Swagelok™-type nut and ferrule
set, and then connect it to the
UHP Nitrogen
inlet until
fingertight. On the other end, connect an appropriate
fitting that connects to your UHP nitrogen gas supply.
Note
Use an open wrench to tighten the fitting on the
gas inlet another 1/8 to 1/4-turn.
(HP nitrogen gas) Cut a length of the
Teflon™ PFA tubing and push it into the
Nitrogen In
inlet. On the other end,
connect an appropriate fitting that
connects to your HP nitrogen gas supply.
1
2
(Orbitrap Fusion Lumos MS)
Содержание Orbitrap Fusion Lumos MS
Страница 1: ...Orbitrap Fusion Series Getting Connected Guide 80011 97002 Revision C June 2017...
Страница 6: ......
Страница 8: ...Contents viii Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...
Страница 24: ...4 Syringe Pump and Modular Valve Assembly 12 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...
Страница 28: ...5 Electrical Connections 16 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...