Thermo Scientific
Orbitrap Fusion Series Getting Connected Guide
11
4
Syringe Pump and Modular Valve Assembly
To attach the syringe pump and the modular divert/inject valve, follow this procedure. The
only tool required is a #2 Phillips screwdriver.
To assemble the syringe pump and modular valve
Note
For instructions on how to set up the syringe pump and plumb the modular valve,
refer to the
Orbitrap Fusion Series Getting Started Guide
.
Using the screwdriver and two of the provided
screws, attach the bottom shelf to the holder.
1
(Not shown) Place the assembly on the top, left
surface of the MS and adjust its position as needed.
4
Center the bottom of the syringe pump on the top
angled, support bracket. Then, using your fingers or
the screwdriver, tighten the knurled, spring-type
screw into the pump.
3
Place the valve on the bottom shelf; using the
screwdriver and two screws, attach the top shelf.
2