6
External Devices
Communication Connectors
Thermo Scientific
Orbitrap Fusion Series Getting Connected Guide
19
4
Start In
Receives the start signal from the contact closure
connection of a connected external device.
To activate this signal, the external device must pull the
signal either low (below 0.75 Vdc) or high (above
2.4 Vdc), depending on the polarity, for at least
100 ms by using a relay, an open-collector driver, or a
similar device that connects between pins 4 and 1.
Note
In the Instrument Configuration window, set
the contact closure signal to “High-to-low edge” or
“Low-to-high edge,” whichever matches the setting
for the connected external device.
5
Ready Out
Provides a relay-driven programmable output signal to
the connected external device. The relay opens when a
method starts and closes when the method finishes.
Output: Maximum 24 Vdc, 3 A
6
Injection Hold
Provides a relay-driven programmable output signal to
the connected external device, such as a fraction
collector.
Output: Maximum 24 Vdc, 3 A
8
RO/IH
Common (return) connection for the Ready Out and
Injection Hold pins.
Analog Input
The two analog channels connect to two separate 12-bit analog-to-digital converters (ADC)
for on-demand conversion of the input voltage. The conversion rate depends on the MS
rate.
1
Chassis
Earth ground (for pins 3 and 4).
3, 4
2V Max:
+ (positive, pin 3) and
– (negative, pin 4)
(Channel 2) Provides a connection for an external
device, such as an LC instrument.
Input: 0–2 Vdc (voltage clamps at 5 Vdc)
5
Chassis
Earth ground (for pins 7 and 8).
7, 8
10V Max:
+ (positive, pin 7) and
– (negative, pin 8)
(Channel 1) Provides a connection for an external
device, such as an LC instrument.
Input: 0–10 Vdc (voltage clamps at 15 Vdc)
Table 3.
Pin-out descriptions for the communication connectors (Sheet 2 of 3)
Pin
Name
Description
Содержание Orbitrap Fusion Lumos MS
Страница 1: ...Orbitrap Fusion Series Getting Connected Guide 80011 97002 Revision C June 2017...
Страница 6: ......
Страница 8: ...Contents viii Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...
Страница 24: ...4 Syringe Pump and Modular Valve Assembly 12 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...
Страница 28: ...5 Electrical Connections 16 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...