Example layout of μDrop Duo Plate containing only blanks
2. Protocol
All photometric measurement types can be used with the μDrop and μDrop Duo Plates. Note that
long kinetic measurements or incubation are not recommended due to evaporation.
The μDrop and μDrop Duo Plates can be used in the whole wavelength range of the instrument in
question.
Figure 9.
Protocol example.
Note
The μDrop and μDrop Duo Plates support only photometric measurements.
Note
The μDrop and μDrop Duo Plates do not support dispensing.
Note
Shaking the μDrop and μDrop Duo Plates is not recommended.