Installation
Requirements
22 Thermo Scientific KingFisher Flex User Manual
Thermo Fisher Scientific
Retain the original packaging and packing material for future transportation.
The packaging is designed to assure safe transport and minimize transit
damage. Use of alternative packaging materials may invalidate the
warranty. Also retain all instrument-related documentation provided by the
manufacturer for future use.
Check the enclosed packing list against order. In case of any deviations,
contact your local Thermo Fisher Scientific representative or Thermo Fisher
Scientific Oy.
Visually inspect the transport package, the instrument and the accessories
for any possible transport damage.
If the carton has been damaged in transit, it is particularly important that
you retain it for inspection by the carrier in case there has also been
damage to the instrument.
Neither the manufacturer nor its agents can be held responsible for any
damage incurred in transit, but the manufacturer will make every effort to
help obtain restitution from the carrier. Upon receipt of the carrier's
inspection report, arrangements will be made for repair or replacement.
If any parts are damaged, contact your local Thermo Fisher Scientific
representative or Thermo Fisher Scientific Oy.
When you set up your KingFisher Flex, avoid sites of operation with excess
dust, vibrations, strong magnetic fields, direct sunlight or UV light, draft,
excessive moisture or large temperature fluctuations.
Make sure the working area is flat, dry, clean and vibration-proof and
leave additional room for accessories, cables, and reagent bottles.
Make sure there is sufficient room behind the instrument to enable
disconnecting the device.
Make sure the ambient air is clean and free of corrosive vapors, smoke
and dust.
Make sure the ambient temperature range is b5°C (41°F) and
+40°C (104°F).
Make sure relative humidity is between 10% and 80% (non-
condensing).
Leave sufficient space (at least 10 cm) on both sides and at the back of the
unit to allow adequate air circulation.
The KingFisher Flex does not produce operating noise at a level that would
be harmful. No sound level measurements are required after installation.
Caution
Do not operate the instrument in an environment where
potentially damaging liquids or gases are present.
s
Checking delivery
Requirements
Содержание KingFisher Flex
Страница 1: ...Thermo Scientific KingFisher Flex User Manual Rev 1 5 Cat no N07669 ...
Страница 88: ...Certificate of Decontamination 88 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Страница 90: ...Glossary 90 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Страница 93: ...Thermo Fisher Scientific Thermo Scientific KingFisher Flex User Manual 93 Notes ...
Страница 94: ...94 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific Notes ...