Installation
How to pack the instrument for transportation
Thermo Fisher Scientific
Thermo Scientific KingFisher Flex User Manual 35
It is recommended that you carry out a check run using a demo protocol to
verify proper instrument operation. Run the check protocol
(Check_96dw_tip, Check_KF96_tip, Check_pcr_tip, or Check_24dw_tip)
under the Maintenance menu according to the KingFisher Flex head and
plastic consumables you are using. If the check is all right, proceed with
your own runs.
This section describes how to refit both the transport locks, the transport
lock of the heating block and that of the tip comb holder.
To refit the transport lock of the heating block, follow these steps:
1. Remove the heating block in the following way:
First choose the Change_heatblock protocol.
Then press
START
.
Remove the heating block by pulling it out.
Press the
START
key when the task has been completed.
2. Switch off the power.
3. Take the transport lock screw from its storage site.
4. Insert the transport lock screw into the label and refit the transport lock
of the heating block into its place (
Figure 3
–26.
Refitting the transport lock of the heating block (A)
How to pack the
instrument for
transportation
How to refit the
transport lock of the
heating block
Содержание KingFisher Flex
Страница 1: ...Thermo Scientific KingFisher Flex User Manual Rev 1 5 Cat no N07669 ...
Страница 88: ...Certificate of Decontamination 88 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Страница 90: ...Glossary 90 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Страница 93: ...Thermo Fisher Scientific Thermo Scientific KingFisher Flex User Manual 93 Notes ...
Страница 94: ...94 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific Notes ...