Installation
Precautions and limitations
Thermo Fisher Scientific
Thermo Scientific KingFisher Flex User Manual 23
Place the instrument on a normal laboratory bench. The net weight of the
entire equipment is approx. 28 kg [62 lbs.].
The instrument operates at voltages of 100
–240 Vac and the frequency
range of 50/60 Hz.
Always ensure that the local supply voltage in the laboratory conforms
to that specified on the type label on the back of the instrument (Figure
Do not smoke, eat or drink while using the KingFisher Flex.
Wash your hands thoroughly after handling test fluids.
Observe normal laboratory procedures for handling potentially
dangerous samples.
Wear proper protection clothing, such as disposable gloves and
laboratory coats, according to good laboratory practice.
Ensure that the working area is well ventilated.
Never spill fluids in or on the equipment.
Caution
The KingFisher Flex should not be kept in close proximity to
magnetic tapes, computer discs or other magnetic storage systems, such
as credit cards, as these can be damaged by the strong magnetic field of
the KingFisher Flex heads.
Do not hold the KingFisher Flex heads close to a PC display, since this
may cause damage to the display.
Do not use metal tools when handling KingFisher Flex heads.
s
Warning
This product contains very strong permanent magnets. People
wearing a pacemaker or metallic prostheses should not use this product. A
pacemaker or prostheses may be affected or damaged if it comes in close
contact with a strong magnetic field.
s
This section describes the installation setups that you must carry out before
operating or relocating the instrument.
The instrument comes with two transport locks (Figure 3
–9): the tip comb
holder transport lock and the heating block transport lock. Remove both the
transport locks.
Make sure both the transport locks are released before you put the
instrument into operation.
Precautions and
limitations
Installation
setups
Releasing the
transport locks
Содержание KingFisher Flex
Страница 1: ...Thermo Scientific KingFisher Flex User Manual Rev 1 5 Cat no N07669 ...
Страница 88: ...Certificate of Decontamination 88 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Страница 90: ...Glossary 90 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Страница 93: ...Thermo Fisher Scientific Thermo Scientific KingFisher Flex User Manual 93 Notes ...
Страница 94: ...94 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific Notes ...